pressure measuring transducer
Specification: |
|
measuring range (MB): overload limit (ÜL): burst pressure (BD): |
1 1.6 2.5
4 6 10 16
25 40 60 100
160 250 400
600 2 3.2 5 8 12 20 32 50 80 120 200 320 500 800 1.200 5 10 10 17 34 34 100 100 400 550 800 1,000 1,200 1,700 2,400 |
|
| output signal: | 4 to 20mA, 2-wire, RA (Ohm) <
(UV(V) - 8V) / 0.02A 0 to 10V, 3-wire, RA > 10 kOhm (other output signals upon request) |
|
| auxiliary energy: | 8 to 30VDC (for output 4 to 20 mA) 14 to 30VDC (for output 0 to 10V) |
|
| accuracy: * | < 1.0% FS (optional. <
0.5% FS)
(*=non-linearity, hysteresis, zero point and scale error incl.. corresponding with error of measurement per IEC 61298-2 Sensor adjusted in vertical mounting position with lower pressure connection) |
|
| non-linearity: | <0.6% FS (optional <0.3% FS) | |
| zero offset: | <0.5% FS (typ.), <0.8% FS (max.) optional: <0.15% FS (typ.), <0.4% FS (max.) |
|
| hysteresis: | <0.16% FS | |
| repeatability: | <0.1% FS | |
| long-termed drift: | <0.1% FS (according to IEC 61298-3) | |
| response time: t90 | <4 ms | |
| perm. temperature of meas.media: | 0°C to +80°C (optionally -30°C to +85°C) | |
| ambient temperature: | 0°C to +80°C (optionally -20°C to +80°C) | |
| storage temperature: | -20°C to +80°C | |
| temperature compensated area: | 0°C to +80°C | |
| temperature error in comp. area: | <1.0% FS (typ.), < 2.5% FS (typ) | |
| material: | Parts coming into contact
with pres. media - pressure connection: 316 L - pressure sensor: 316 L (as of 10bar rel. 13 to 8 PH) - housing: 316 L |
|
| pressure connection: | G 1/4A (= 0,635 cm), DIN 3852-E with NBR sealing | |
| protection rating: | IP65 resp. IP67 with cable | |
| electric connection: | elbow-type plug acc. to DIN 43650 or connection cable, cable length 1.5m | |
| electric protection: | reverse voltage and short-circuit protection | |
| weight: | approx. 80g | |
Options,Accessories (against upcharge): |
| absolute pressure (0 to 1 bar abs. to 0 to 25 bar abs.) |
| Higher sensor accuracy (class 0.5) upcharge: |
| Extended temperature range upcharge: |
| Output signal 0 to 10V upcharge: |
| Fixed connecting cable,
2m with bend protection upcharge: (instead of elbow-type plug, IP67) |
| GWA 1214 V4A thread adapter G 1/2" (= 1,27 cm) with international thread G1/4" (= 0,635 cm)and external thread G1/2" (= 1,27 cm) |
(c) 1999-2010 GSG Geologie-Service GmbH, Wuerzburg -
www.geologie-service.com
Phone +49/931 99105-80; FAX: +49/931 99105-90
- All rights reserved -