specifications: |
|
|
measuring input |
display range, solution |
accuracy |
measuring rate |
|
|
GIR 230 NS |
4 to 20mA, 0 to 20mA, 0 to 10V |
-1999 to 9999 digit, scale freely adjustable recommende range <2000 digit |
<0.2% f.s. +1 digit
|
approx. 100 measurings/sec. |
|
|
GIR 230 Pt |
Pt100 (3-wire), Pt1000 (2-wire) |
Pt100: |
-200°C to +850°C resp. -50°C to +200°C |
<0.5% f.s. +1 digit
|
approx. 4 measuringsq/sec. |
| Pt1000: | -200°C to +850°C | ||||
|
GIR 230 TC |
types J, K, N, S, T and 0 to 50mV |
type J: |
-170°C to +950°C |
<0.3% f.s. +1 digit
(type S: <0.5% f.s. +1 digit) point of comparison: +1°C |
approx. 4 measurings/sec. |
| type K: | -270°C to +1350°C | ||||
| type N: | -270°C to +1300 °C | ||||
| type S: | -50°C to +1750°C | ||||
| type T: | -270°C to +400°C | ||||
|
GIR 230 FR |
frequency (TTL-signal) |
-1999 to 9999 digits, freely adjustable |
<0.2% f.s. +1 digit
|
|
|
| frequency meas.: | 0,000 Hz to 10 kHz | ||||
| rotational speed: |
0,000U/min to 9999 U/min, selectable prescaler (1-1000) |
||||
| Counter up/down: |
0 to 9999 (~10.000.000 with prescaler) |
||||
|
GIR 230 NTC |
NTC (2-wire) |
-40.0°C to +120.0°C |
<0.5% f.s. +1 digit
|
approx. 4 measurings/sec. |
|
|
GIR 230 Pt1000/DIF, |
2x Pt1000 (2-wire) |
Pt1000: |
-200°C to +850°C, 1°C |
<0.5% f.s. +1 digit
|
approx. 4 measurings/sec. |
| NTC: | -40.0°C to +120.0°C, 0.1°C | ||||
| display: | difference temperature sensor1 - sensor2 | ||||
|
GIR 230 NS/DIF |
(2x) 4 to 20mA, (2x) 0 to
20mA Please specify the input signal wished in your order! |
-1999 to 9999 digits,
initial, end values and DPs |
<0.2% f.s.+1 digit
|
approx. 100 measurings/sec. |
|
| outputs: | 2 relay outputs, 230V~ switching (GIR 230 NTC: 1 relay output) | |
| relay output: | 2 (1) closing contacts, switching power 5A, 230V AC | |
| alarm output: | NPN, open collector, switching power: 30mA, max. 28V | |
| controller states: | 2-point, 3-point*, 2-point with alarm, min-/max-alarm to 1 output, min-/max-alarm to 2 outputs* (*= not available at GIR 230 NTC) |
|
| switching points, hysteresis, alarm points: | freely selectable | |
| min-/max-value memory: |
the max- and min-value will be stored |
|
| display: |
approx. 10mm high, 4-digit red LED-display |
|
| operation: | via 3 buttons | |
| operating condtitions: | -20°C to +50°C, 0 to +80% r.h. (non condensing) | |
| power supply: | 230V, 50/60Hz, approx. 2VA | |
| housing: | glass fibre reinforced Noryl, front panel PC | |
| dimensions. | 24 x 48mm (front frame) | |
| mounting depth: | approx. 65mm (incl. screw-type/plug-in terminal) | |
| panel mounting: | with VA-spring clamp | |
| allowed panel thicknesses from 1 to aprox. 10mm | ||
| panel cut-out: | 21.7+0.5 x 45+0.5mm (H x W) | |
| connection terminal: | screw-type/plug-in terminal | |
| 4-pin (...NTC: 3-pin) for power supply and relay outputs | ||
| 4-pin (...NTC: 3-pin) for measuring input and alarm output | ||
| for wire cross selections from 0.14 to 1.5mm2 | ||
| IP rating: | front side IP54 (IP65 upon request) | |
| Noise immunity (EMC): | EN61326 (appendix A, class B) | |
suitable temperature probes (against upcharge): |
| temperature probes (Pt100/Pt1000) | |
| temperature probes (type K, S, N) | |
| GTF 230S |
NTC-temperature probe -40°C to +120°C sensor sleeve made of st. steel, diameter 5 x 50mm, cable made of silicone, cable length aprox. 1m |
| Options: | with longer probe cable (upcharge each m) |
Accessories (against upcharge): |
| GGD 2448 SET: | O-rings for device mounting IP65 (2 pieces) |
(c) 1999-2010 GSG Geologie-Service GmbH, Wuerzburg -
www.geologie-service.com
Tel. +49/(0)931
30 40 8-0; FAX: +49/(0)931 99105-90
- All rights reserved -