Temperature transmitter with digital adjustment

 

GTMU-IF1                      

standard: probe length FL = 100mm, D = 6mm

GTMU-IF2                      

standard: probe length FL = 100mm, D= 6mm G = 1/2" (= 1,27 cm)

GTMU-IF3                      

standard: probe length FL= 50mm, HL = 100mm, D= 6mm, G = 1/2" (= 1,27 cm)

 

Specification:

 
Measuring range: GTMU-IF1: -30 to +100°C (standard)
GTMU-IF2:
-30 to +100°C (standard)
GTMU-IF3:
-70 to +400°C (standard)

The probe length "FL" has to be chosen long enough,
that the allowable temperature range of the electronics
situated in the tube sleeve is not exceeded.

other measuring ranges (max. -200 to +500°C) upon request
 

Accuracy: (At nominal temperature = 25°C)
Electronic: +0,2% of meas. value +0,2 °C
Measuring probes: standard: DIN class B (optionally higher sensor accuracy available)
 
Measuring probe: internal Pt1000-sensor
Output signal: standard: 4 - 20 mA (2-wire),
optionaly: 0-1V, 0-10 V (other signals upon request)
Auxiliary energy: Uv  = 10 ... 30 V DC
Permissible burden: RA <(UV - 10V)/0,022 A RA in Ohm, UV in V)
Scaling: The transducer can be scaled freely within the
measuring ranges via GTMU-IF programming tool
Operating temperature of electronic: (in tube sleeve): -25 to  + 60 °C
Electric connection: approx. 1m long 4-pin cable
(2 x current loop, 2 x interface)
Thread: Standard: G 1/2" (= 1,27 cm)
Optionally: G 1/8" (= 0,3175 cm), G 1/4" (= 0,635 cm), G 3/8" (= 0,9525 cm), G 3/4" (= 1,905 cm), M8x1, M10x1, M14x1.5
Housing: Stainless steel housing
Dimensions: depending on sensor construction
Tube: stainless
Tube diameter "D": 3, 4, 5, 6 oder 8 mm (6mm = standard)
Tube length "FL": 100 or 50 mm or on customer requirement
Collar tube length "HL": 100 mm or on customer requirement
Tube sleeve: diameter 15x35 mm (without screwing)

The probe length "FL" has to be chosen long enough, that the allowable
temperature range of the electronics situared and of the housing not overstop 70°C!

 

Options (upcharges):

- FL Longer tube, each started further 100mm
-HL Longer collar tube, each started further 100mm
-D other tube diameter
-G other thread
-MB other measuring ranges, set by factory

 

Accessories (against upcharge):

GTMU-IF-Programming tool:  USB-interface adaptor for GTMU-IF, incl. programming software

(c) 1999-2010 GSG Geologie-Service GmbH, Wuerzburg  -  www.geologie-service.com
Tel. +49/(0)931 30 40 8-0; FAX: +49/(0)931 99105-90